Z-1329-2025: Z1 Femoral Hip System, Product Number 611777612

Recall record Z-1329-2025 for Zimmer, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1329-2025
Company in FDA record
Zimmer, Inc.
FDA classification date
Mar 24, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a potential design issue with the Z1 Offset Stem Inserter leading to fracturing of the distal tip.

Source facts

Action
An URGENT MEDICAL DEVICE RECALL notification dated 2/12/25 was sent to customers. Our records indicate that you may have received one or more of the affected products. The affected units were distributed between September 2024 and January 2025. Risk Manager Responsibilities: 1. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3. If the product has been further distributed, provide your customers with the recall notice and ensure documentation. 4. Complete Attachment 1 Certificate of Acknowledgement and send to CorporateQuality.PostMarket@zimmerbiomet.com. This form shall be returned even if you do not have affected products at your facility. 5. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facility s documentation. 6. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to CorporateQuality.PostMarket@zimmerbiomet.com. Surgeon Responsibilities: 1. Review this notification for awareness of the contents. 2. For cases where the Z1 Offset Stem Inserter was used, increased clinical or radiographic monitoring may be required if the device fractured during use. If the Z1 Offset Stem Inserter was not used or if there was no evidence of fracture during use, there are no additional specific patient monitoring instructions related to this recall that are recommended beyond your existing follow-up schedule. 3. Complete Attachment 1
Initiated
2025-02-12
Root cause
Under Investigation by firm
Distribution
US Nationwide distribution.
Recall number
Z-1329-2025
Recall status
Open, Classified
CDRH record ID
212606
FDA classified
2025-03-24
Recall event number
96348

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗