Reason stated in the source
The Liquichek" ToRCH Plus IgM Control - Positive, contains detectable levels of HCV RNA which originated from the non-specific human IgM component used in the mouse monoclonal antibody conjugation process for this product.
Source facts
- Action
- Bio-rad Laboratories initiated this recall on April 18, 2013, via telephone call and certified mailing to customers. The notification letter titled "Important Product Information Notice of Correction" was accompanied by a "Field Correction Notification Acknowledgement" form that notified customers of the product for recall (description and codes), issue (This product contains detectable levels of HCV RNA which originated from the non-specific human IgM component used in the mouse monoclonal antibody conjugation process for this product), Customers were instructed to immediately discontinue use this lot of Liquichek ToRCH Plus IgM Control. Discard all remaining inventory of this lot of Liquichek ToRCH Plus IgM Control that you have at your location in accordance with the requirements of your local waste management authorities. Handle this product with the same precautions used with patient specimens. In accordance with good laboratory practice, all human source material should be considered potentially infectious. Have the appropriate laboratory personnel complete the acknowledgment form and email or fax it to us. It is very important to us to know that you have received the information provided in this Notice. Contact information (1-800-2-BIORAD)
- Initiated
- 2013-04-15
- Root cause
- Material/Component Contamination
- Terminated
- 2013-08-01
- Distribution
- Worldwide Distribution - Distributed in San Juan Puerto Rico and the states of NM, NY, MN, OR and WI., and the countries of Australia, Brazil, France, Italy, Korea, and New Zealand.
- Recall number
- Z-1330-2013
- Recall status
- Terminated
- CDRH record ID
- 117531
- FDA classified
- 2013-05-16
- Recall event number
- 64959