Skip to content

Public FDA record

Z-1330-2013: Liquichek" ToRCH Plus IgM Control - Positive Liquichek ToRCH Plus IgM Control is intended for use as an

Recall record Z-1330-2013 for Bio-rad Laboratories, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1330-2013
Company in FDA record
Bio-rad Laboratories, Inc.
FDA classification date
May 16, 2013
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class III

Reason stated in the source

The Liquichek" ToRCH Plus IgM Control - Positive, contains detectable levels of HCV RNA which originated from the non-specific human IgM component used in the mouse monoclonal antibody conjugation process for this product.

Source facts

Action
Bio-rad Laboratories initiated this recall on April 18, 2013, via telephone call and certified mailing to customers. The notification letter titled "Important Product Information Notice of Correction" was accompanied by a "Field Correction Notification Acknowledgement" form that notified customers of the product for recall (description and codes), issue (This product contains detectable levels of HCV RNA which originated from the non-specific human IgM component used in the mouse monoclonal antibody conjugation process for this product), Customers were instructed to immediately discontinue use this lot of Liquichek ToRCH Plus IgM Control. Discard all remaining inventory of this lot of Liquichek ToRCH Plus IgM Control that you have at your location in accordance with the requirements of your local waste management authorities. Handle this product with the same precautions used with patient specimens. In accordance with good laboratory practice, all human source material should be considered potentially infectious. Have the appropriate laboratory personnel complete the acknowledgment form and email or fax it to us. It is very important to us to know that you have received the information provided in this Notice. Contact information (1-800-2-BIORAD)
Initiated
2013-04-15
Root cause
Material/Component Contamination
Terminated
2013-08-01
Distribution
Worldwide Distribution - Distributed in San Juan Puerto Rico and the states of NM, NY, MN, OR and WI., and the countries of Australia, Brazil, France, Italy, Korea, and New Zealand.
Recall number
Z-1330-2013
Recall status
Terminated
CDRH record ID
117531
FDA classified
2013-05-16
Recall event number
64959

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗