The DP-30 displays an incorrect needle-guide bracket when used with the 65EC10EA model
transducer.
Source facts
Action
On January 13, 2020, the firm distributed Urgent Medical Device Correction letters to affected distributors. Distributors were informed that the system displays the incorrect needle-guide bracket when used with the 65EC10EA model transducer.
Customers were advised to avoid use of NGB-004 needle-guide with transducer model 65EC10EA until a software update has been applied to the system. The system may be used normally for all other functions. The firm recommended that distributors immediately notify any clients to which they have distributed affected systems (DP-30 systems with software version 03.00.00 and 03.01.00 distributed between 06/25/2018
and 06/28/2019), or provided a software update (USB with V03.01.00, part number 110-002261-05).
Within the next several weeks, Mindray will provide a software update at no cost. Should you
require technical assistance while performing the update, please contact Mindrays Technical
Support team at 800-288-2121, option 2, Monday through Friday, 8:30am - 5:30pm PT.
The replacement USB drive contains instructions on how to perform the software update, under
the Manuals folder in the DP-10 & DP-20 & DP-30 Series System Recovery Guide file.
If you have questions regarding this recall, you may contact Diane Arpino, Director, Quality Operations and Regulatory Affairs, at 201-995-8407, or via email at d.arpino@mindray.com.
Initiated
2020-01-13
Root cause
Software Design Change
Terminated
2020-08-21
Distribution
US Nationwide distribution in the states of CA, GA, IL, MD, NY, SC, and WA.
Recall number
Z-1330-2020
Recall status
Terminated
CDRH record ID
179418
FDA classified
2020-02-05
Recall event number
84794
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.