Z-1330-2021: Product Label - ACUSON Sequoia *** Power Input: 100-240V~, 13.0-5.4A, 50-60Hz *** (240)11148775 Model Us

Recall record Z-1330-2021 for Siemens Medical Solutions USA, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1330-2021
Company in FDA record
Siemens Medical Solutions USA, Inc.
FDA record creation date
Apr 1, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The ultrasound system averages the Mean Sac Diameter (MSD) and Gestational Sac Diameter (GSD) into the Estimated Date of Delivery (EDD) calculation. This may result in an incorrect EDD, which may influence patient management decisions regarding induction of labor and elective caesarean delivery, which may result in premature births.

Source facts

Action
On 02/04/2021, the firm sent an "CUSTOMER SAFETY ADVISORY NOTIFICATION" email to customer informing them that the ultrasound system averages the Mean Sac Diameter (MSD) and Gestational Sac Diameter (GSD) into the Estimated Date of Delivery (EDD) calculation. This may result in an incorrect EDD. An incorrect EDD may influence patient management decisions regarding induction of labor and elective caesarean delivery, which may result in premature births. -Customers are instructed to avoid this issue, delete any MSD or GSD values from the patient report if other biometric parameters are measured. -Recalling Firm recommends a review of any early obstetric or gynecology ultrasound results for current pregnancies which produced a gestational age using MSD or GSD as an input. -The issue will be corrected by the Recalling Firm with an updated to the software and a Customer Service Engineer will contact customers to schedule a facility visit to update the software or updates will be issued via the Remote Update Handling process. -Customers are instructed to inform their personnel accordingly. Additional questions or assistance - contact Quality & Regulatory at 732-331-6672 or email khalil.thomas@siemens-healthineers.com.
Initiated
2021-02-04
Root cause
Software Design Change
Distribution
U.S.: AK, AL, AR, AZ, CA, CT, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY O.U.S.: U.A.E., Algeria, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hong Kong, India, Indonesia, Iran, Israel, Italy, Japan, Kazakhstan, Kuwait, Mexico, Morocco, Netherlands, New Zealand, Oman, Philippines, Poland, Portugal, Qatar, Republic Korea, Romania, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Ukraine, United Kingdom, and Vietnam
Recall number
Z-1330-2021
Recall status
Open, Classified
CDRH record ID
185877
FDA record created
2021-04-01
Recall event number
87362

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗