Skip to content

Public FDA record

Z-1330-2024: Situate Laparotomy Sponge, RF and X-ray Detectable, REF: L1818-04P01C-1

Recall record Z-1330-2024 for Covidien. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1330-2024
Company in FDA record
Covidien
FDA classification date
Mar 21, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Packs including 5 laparotomy sponges may contain incorrect product with a blue loop, X-ray detectable feature, and ethylene oxide treatment; however, this incorrect product wasn't steam pre-treated prior to sterilization, and has no RF tag, which if non-sterile, may lead to infection, tissue trauma, and sepsis, and if in the scannable area, would lead a "CLEAR" display.

Source facts

Action
On 2/21/24, recall notices were mailed to customers and distributors who were asked to do the following: 1) Do not use impacted products. 2) Remove, quarantine, and return all unused affected product. 3) This communication needs to be passed to all those who need to be aware within your organization where impacted devices have been transferred, and to any other associated organizations that may be impacted by this action. 4) Complete and return the customer confirmation form to rs.gmbmitgfca@medtronic.com Customers with questions can contact the firm at 800-962-9888 Option 2.
Initiated
2024-02-21
Root cause
Labeling mix-ups
Distribution
US Nationwide distribution including in the states of AK, AL, AZ, CA, CO, FL, GA, IA, IL, IN, KS, LA, MA, MD, MI, MN, MO, MS, NC, NE, NJ, NM, NY, OH, OK, PA, RI, TN, TX, UT, VA, WA, WI, WV.
Recall number
Z-1330-2024
Recall status
Open, Classified
CDRH record ID
206389
FDA classified
2024-03-21
Recall event number
94127

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗