Open original FDA record ↗Recall record
- FDA identifier
- Z-1330-2024
- Company in FDA record
- Covidien
- FDA classification date
- Mar 21, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Packs including 5 laparotomy sponges may contain incorrect product with a blue loop, X-ray detectable feature, and ethylene oxide treatment; however, this incorrect product wasn't steam pre-treated prior to sterilization, and has no RF tag, which if non-sterile, may lead to infection, tissue trauma, and sepsis, and if in the scannable area, would lead a "CLEAR" display.
Source facts
- Action
- On 2/21/24, recall notices were mailed to customers and distributors who were asked to do the following: 1) Do not use impacted products. 2) Remove, quarantine, and return all unused affected product. 3) This communication needs to be passed to all those who need to be aware within your organization where impacted devices have been transferred, and to any other associated organizations that may be impacted by this action. 4) Complete and return the customer confirmation form to rs.gmbmitgfca@medtronic.com Customers with questions can contact the firm at 800-962-9888 Option 2.
- Initiated
- 2024-02-21
- Root cause
- Labeling mix-ups
- Distribution
- US Nationwide distribution including in the states of AK, AL, AZ, CA, CO, FL, GA, IA, IL, IN, KS, LA, MA, MD, MI, MN, MO, MS, NC, NE, NJ, NM, NY, OH, OK, PA, RI, TN, TX, UT, VA, WA, WI, WV.
- Recall number
- Z-1330-2024
- Recall status
- Open, Classified
- CDRH record ID
- 206389
- FDA classified
- 2024-03-21
- Recall event number
- 94127