Z-1331-2021: Alaris Infusion Pump Module, Model 8100, serviced/repaired by Tenacor with an impacted bezel repair part

Recall record Z-1331-2021 for Tenacore LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1331-2021
Company in FDA record
Tenacore LLC
FDA classification date
Apr 2, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Bezel repair parts used, not by the original manufacturer, to service and repair infusion pump modules, may over time experience cracking or separation of the posts from the bezel. The separation of one or more bezel repair part posts may result in free flow, over infusion, under infusion or interruption of infusion.

Source facts

Action
On 02/25/2021, Field Safety Corrective Action notifications were mailed to customers; customers were also notified via phone. The recalling firm provided the following services to customers: serviced infusion pumps, sold infusion pumps, and sold bezel repair parts. Customers were asked to do the following: 1. Review the serialized list of potentially impacted pumps and remove these devices from clinical use and immediately quarantine them until they can be inspected, and impacted parts are replaced, if necessary. Part customers were told to quarantine affected parts. 2. Inspect the potentially affected devices to determine if the impacted bezel repair part is installed. Refer to inspection instructions and a) identify the impacted Alaris 8100 Infusion Pump Module by serial number; b) Look from the top down for the ejector pin marks, c) If NO ejector pin marks are identified, this module requires bezel. Customers who were sold pumps can identify affected pumps through serial number alone. 3. Dealers and Distributors only: If potentially impacted pumps were sold, immediately identify, inspect and recall, if necessary, the potentially affected pumps. An electronic copy of a proposed field safety corrective action notice will be provided that can to be used for the facilities that may be using the potentially affected pumps. 4. Contact the recalling firm to schedule bezel repair part replacement. The impacted bezel repair part must be replaced before the pump can be returned to service. 5. Please immediately share this field safety corrective action notification with all necessary parties within your organization to ensure that they are also aware of this action. Notify end user customers. 6. Complete and return the customer response form. Contact Quality Assurance at 714-371-9919 or Quality@Tenacore.com to schedule replacement of the impacted part.
Initiated
2021-02-25
Root cause
Component design/selection
Terminated
2024-08-05
Distribution
US distribution to states of: WA, AZ, NY, TX, LA, CA, NV, ME, MN, DE, NC, FL, MO, KY, TN, and PA
Recall number
Z-1331-2021
Recall status
Terminated
CDRH record ID
186087
FDA classified
2021-04-02
Recall event number
87470

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗