D.O.R.C. Dutch Opthalmic Research Center Intl B.V.
FDA classification date
Mar 21, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II
Reason stated in the source
When using those products you may experience difficulties to extend or retract the laser fiber, and in some occasions may also have difficulties to direct the
laser fiber tip precisely.
Source facts
Action
An Urgent Medical Device Recall notification letter was sent to customers on 01/15/2024. Actions to be Taken by the Customer: Pass this letter on to all those who need to be aware within your organization and/or to any organization where the potentially affected devices have been transferred to. Verify whether you have any unopened boxes, or individual pouches of the affected DORC directional laser probes in your inventory. Remove any remaining boxes and individual Products from your inventory and return the impacted Product to DORC following the instructions in the attachment. Complete, even if you do not return the Product, the on-line reply form per instructions in the attachment. In case of questions please contact us in any of the following manners: Contact your local DORC representative, or call the Customer Service at our distributor Dutch Ophthalmic, USA at 1-800-753-8824 x 1, or send an email to TSC@dorcglobal.com.
Initiated
2024-01-15
Root cause
Process control
Distribution
US Nationwide distribution in the states of AZ, CO, FL, IN, MD, NC, NY, OH, SC, TX and PR.
Recall number
Z-1331-2024
Recall status
Open, Classified
CDRH record ID
206145
FDA classified
2024-03-21
Recall event number
94022