Open original FDA record ↗Recall record
- FDA identifier
- Z-1333-04
- Company in FDA record
- Welch Allyn Protocol, Inc
- FDA classification date
- Aug 10, 2004
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Potential speaker failure could result in absence of audible annunciation of pulse tones, patient alarms and equipment alerts.
Source facts
- Action
- The firm contacted their customers and their distributors customers via letters dated June 8, 2004 through July 9, 2004. The letters instruct their customers to verify speaker functionality. If the customers do not hear two quick beeps or there is static then they have a faulty speaker. The letters instruct them to stop using the monitor and contact Welch Allyn to have the monitor repaired. If the monitor continues to emit two quick beeps their customers are to contact Welch Allyn to have the monitor repaired as soon as it is convenient. The customers are to verify the two beeps prior to each use. The letter dated June 8, 2004 went to only a couple of direct customers. The remaining letters dated June 14, 2004, June 23, 2004, June 30, 2004 and July 9, 2004 went to direct customers, distributors and distributors'' end user customers.
- Initiated
- 2004-06-08
- Root cause
- Other
- Terminated
- 2006-06-21
- Distribution
- Welch Allyn distributes product to hospitals, physicians and distributors nationwide and internationally.
- Recall number
- Z-1333-04
- Recall status
- Terminated
- CDRH record ID
- 33369
- FDA classified
- 2004-08-10
- Recall event number
- 29233