Open original FDA record ↗Recall record
- FDA identifier
- Z-1333-2014
- Company in FDA record
- Intuitive Surgical, Inc.
- FDA classification date
- Apr 4, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Intuitive Surgical has become aware of the potential for certain lots of the Single-Site 5mm Curved Cannulae to become damaged during use. The second part of this action is to update Single-Site labeling. These labeling changes instruct users how to inspect cannulae prior to use and to appropriately handle cannulae during use.
Source facts
- Action
- Intuitive sent an Urgent Medical Device letter dated March 25, 2014, to all users of cannulas that have the potential to be damaged during use as well as a second letter to all single site cannula users that provided clarity to instruction for use for inspections and handling of the product. Certain lots manufactured prior to August of 2012, and users were asked to use Attachment A to help identify affected lots numbers. Phase 1 - Cannulas should then be inspected with information provided in Attachment B. Any damaged devices should be quarantined and returned to intuitive. If inspection does not find damage, the device may continue to be used. Any other cannulae and any non damaged product should be used according to Attachment C's instructions for use. Phase 2 - Once replacement cannulae are available, the undamaged cannulae should be returned. The Acknowledgement form should be completed and returned to Intuitive. Intuitive will contact consignees when replacement cannulae are available. Questions should be directed to Customer Service at 800-876-1310. Those consignees that use the 5 mm canulae but do not have affected product were provided with a letter that only includes the Single Site Cannula inspection and Use addendum.
- Initiated
- 2014-03-13
- Root cause
- Device Design
- Terminated
- 2014-09-03
- Distribution
- Worldwide Distribution - USA and the countries of Australia, Belgium, Brazil, Chile, China, Columbia, Denmark, France, Germany, India, Italy, Lebanon, Monaco, Netherlands, Norway, Romania, Saudi Arabia, South Korea, Spain, Sweden, Switzerland, Turkey and United Kingdom.
- Recall number
- Z-1333-2014
- Recall status
- Terminated
- CDRH record ID
- 126238
- FDA classified
- 2014-04-04
- Recall event number
- 67758