Reason stated in the source
Endoscope cleaning brush 11276CL may not fit resulting in an inability to insert the brush in to the channel of the intended endoscope or to completely pass it through such channel.
Source facts
- Action
- Karl Storz issued Recall Letter to users on 12/23/15. Firm's letter: KARL STORZ has identified a replacement brush; model 11276CL2 that is now available to clean your endoscope. Please check your inventory and remove any model 11276CL brushes. These brushes must be returned to KARL STORZ. Please complete the attached form and followthe return instructions. Upon receipt of any 11276CL brushes from you we will provide you the 11276CL2 cleaning brush as replacement at no charge. Contact Lynn Carter, Director Quality Assurance at 508-248-9011, extension 1664 Monday through Friday, 8:00AM to 4:30PM, Eastern Time.
- Initiated
- 2016-02-29
- Root cause
- Component design/selection
- Terminated
- 2016-09-28
- Distribution
- Nationwide Foreign: Canada, Mexico,
- Recall number
- Z-1333-2016
- Recall status
- Terminated