Z-1333-2021: BIOCARE Medical , IntelliPath FLX, Automated Staining Instrument, Ref: " Model Number:IPS0001US and IPS0

Recall record Z-1333-2021 for Biocare Medical, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1333-2021
Company in FDA record
Biocare Medical, LLC
FDA record creation date
Apr 1, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Their is a potential that the automated staining instrument with software version 3.5.3.1 may move with random speeds in random direction across its range of motion. This could result in the instrument hitting end stops or running into other components on the working deck. This could also result in wash buffer being dispensed on random locations.

Source facts

Action
On 2/25/2021, BIOCARE Medical emailed and "Medical Device Recall" Notification to all affected consignees. In addition to informing consignees about the recall, the firm requested consignees take the following actions: 1. Do not use the product in the Multi Batching mode. 2. Do not interrupt a run during a Single Batch mode run or access the working deck while the instrument is in operation. 3. Always run positive controls per CAP and CLIA guidance to avoid the potential for false negative result from contaminated reagents. 4. If the malfunction does occur: " Do not open the cover and access the working deck. " Power cycle (turn off, then on) the device to reset it. " Look for spilled Wash Buffer solution on the lab bench and the floor. " Carefully examine the working deck and to determine if wash buffer has been spilled. If there is evidence of spilled wash buffer near, on or in reagent vials those vials should be disposed of to prevent staining performance issues. 5. PLEASE INDICATE YOUR UNDERSTANDING AND ACKNOWLEGEMENT OF THIS NOTICE BY COMPLETING THE STEPS LISTED BELOW 1. Locate all intelliPATH FLX Automated Staining Instruments in your possession. Verify the software version to confirm your device is impacted by this notice. 2. Ensure all personnel who may utilize the intelliPATH FLX are aware of this recall. 3. Complete and return the attached Response Form to Biocare Medical to acknowledge your receipt and understanding of this Recall Notice within 10 days of receipt. No return of product is necessary. This notification is being provided for awareness only. Biocare Medical continues to investigate this matter and will follow up with affected customers. Adverse events or quality problems experienced with the use of this product must be reported to Biocare Medical via techsupport@biocare.net.
Initiated
2021-02-25
Root cause
Software Design Change
Terminated
2023-01-31
Distribution
US: AL, AZ, CA, CT, FL, IL, MA, MD, MN, NC, OH, OR, PA, WA, WY OUS: Germany, Netherlands
Recall number
Z-1333-2021
Recall status
Terminated
CDRH record ID
186007
FDA record created
2021-04-01
Recall event number
87426

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗