Z-1333-2022: GE Centricity Universal Viewer Zero Footprint. To view, communicate, process, and display Medical image

Recall record Z-1333-2022 for GE Healthcare, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1333-2022
Company in FDA record
GE Healthcare, LLC
FDA classification date
Jul 5, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential to display inaccurate measurements on images in Centricity Universal Viewer Zero Footprint Client (ZFP)

Source facts

Action
GE notified their consignees by issuing an URGENT MEDICAL DEVICE CORRECTION notice on 08 June 2022. The letter explains the two safety issues and provides the following actions be taken: Issue #1: "You can continue to use your device in accordance with the User Manual and the actions below: Do not utilize any measurements when the 'Lossy' overlay is displayed in the lower left of the viewport and, when you hover over the word 'Lossy', it displays the tooltip box with the phrase 'Image scaled down from its original resolution'." Issue #2: "You can continue to use your device in accordance with the User Manuals and the actions below: Do not perform any measurement on non-squared pixel images. Always confirm critical diagnostic measurements by other means (i.e using a fiducial marker of known size)." GE Healthcare will correct all affected products at no cost to you. A GE Healthcare representative will contact you to arrange for the correction. After the software has been corrected, be sure to destroy the installation media for affected software at your site. If you have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
Initiated
2022-06-08
Root cause
Software design
Distribution
Worldwide distribution.
Recall number
Z-1333-2022
Recall status
Open, Classified
CDRH record ID
194437
FDA classified
2022-07-05
Recall event number
90454

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗