Z-1334-04: 8870 software application card versions AAA 02, BBB 04, BBC 02, and BBD 01, which is used in conjunction

Recall record Z-1334-04 for Medtronic Inc, Neurological & Spinal Division. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1334-04
Company in FDA record
Medtronic Inc, Neurological & Spinal Division
FDA classification date
Sep 23, 2004
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

With the Model 8870 software application card versions AAA 02, BBB 04, BBC 02, and BBD 01, which is used in conjunction with the Model 8840 N'Vision Clinician Programmer, users may have mistakenly entered a periodic bolus interval into the minutes field, rather than the hours field, resulting in drug overdoses. This issue is limited to programming the SynchroMed and SynchroMed EL pumps.

Source facts

Action
Letters were distributed 8/24/2004 utilizing a courier with signed receipt upon delivery. The letters were signed for upon receipt at the consignee address, and the contents of the letter included a response form which can be faxed or mailed back to Medtronic. If any letters are returned as 'undeliverable', attempts will be made to telephone these consignees. A press release was issued on September 22, 2004 by Medtronic Inc. for all Version AAA 02 Model 8870 software application cards in the U.S.
Initiated
2004-08-24
Root cause
Other
Terminated
2005-12-13
Distribution
These cards were distributed primarily in the USA, but also to countries in Europe, Middle East and Africa, Canada, Australia, Hong Kong, Latin America and Japan.
Recall number
Z-1334-04
Recall status
Terminated
CDRH record ID
34649
FDA classified
2004-09-23
Recall event number
29849

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗