Open original FDA record ↗Recall record
- FDA identifier
- Z-1334-06
- Company in FDA record
- ABX Diagnostics Inc
- FDA classification date
- Aug 3, 2006
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Labels can be incorrectly placed on a sample tube resulting in a possible error in results attributed to a sample.
Source facts
- Action
- Notifications have been prepared advising that following an occurrence observed outside of the United States, HORIBAABX is providing information regarding the rare circumstance of a label being incorrectly placed on a sample tube, resulting in a possible error in results attributed to a sample. The March 24, 2006, letter identifies various conditions in which the anomaly occurs. In order to prevent this type of incident, the firm recommends that the customer ensure that the identification label is firmly attached to the sample tubes during their positioning or removal from the racks.
- Initiated
- 2006-03-24
- Root cause
- Other
- Terminated
- 2012-05-23
- Distribution
- Nationwide
- Recall number
- Z-1334-06
- Recall status
- Terminated
- CDRH record ID
- 46682
- FDA classified
- 2006-08-03
- Recall event number
- 35719