Reason stated in the source

Labels can be incorrectly placed on a sample tube resulting in a possible error in results attributed to a sample.

Source facts

Action
Notifications have been prepared advising that following an occurrence observed outside of the United States, HORIBAABX is providing information regarding the rare circumstance of a label being incorrectly placed on a sample tube, resulting in a possible error in results attributed to a sample. The March 24, 2006, letter identifies various conditions in which the anomaly occurs. In order to prevent this type of incident, the firm recommends that the customer ensure that the identification label is firmly attached to the sample tubes during their positioning or removal from the racks.
Initiated
2006-03-24
Root cause
Other
Terminated
2012-05-23
Distribution
Nationwide
Recall number
Z-1334-06
Recall status
Terminated
CDRH record ID
46682
FDA classified
2006-08-03
Recall event number
35719

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗