Z-1334-2021: CF Delivery Dev Caps Bravo x5, FGS-0635/FGS 0636 - Product Usage: intended to be used for gastroesophagea

Recall record Z-1334-2021 for Covidien Llc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1334-2021
Company in FDA record
Covidien Llc
FDA record creation date
Apr 1, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The capsule may fail to attach to the esophageal mucosa, which in turn may lead to the aspiration of the capsule. In the event of capsule aspiration, immediate intervention to remove the capsule is required. Potential outcomes following capsule aspiration include low oxygen saturation, intervention to retrieve the capsule, potential need to intubate the patient, extended hospital stay and a delay of treatment. Thirteen reports of capsule aspiration have been received the last 2 years.

Source facts

Action
On February 26, 2021, the firm initiated mailing of Urgent Medical Device Recall letters to affected customers. Customers were advised that in cases where the capsule fails to attach to the esophageal mucosa, the potential exists for aspiration of the capsule. In the event of capsule aspiration, immediate intervention to remove the capsule is required. Potential outcomes following capsule aspiration include low oxygen saturation, intervention to retrieve the capsule, potential need to intubate the patient, extended hospital stay and a delay of treatment. Thirteen reports of capsule aspiration have been received in the last 2 years. Manufacturing process improvements have been implemented to address this issue. Customers were asked to do the following: 1. Please immediately quarantine and discontinue use of affected item codes and lots. 2. Please return affected product as indicated in the letter. All unused products from the affected item codes and lots must be returned. 3. If you have distributed the affected Bravo capsules, please promptly forward the information from this letter to those recipients. 4. Complete the Recalled Product Return Form even if you do not have inventory. If you have any questions regarding this communication, please contact your Medtronic representative or Customer Service at 800-448-3644, option 2, option 2.
Initiated
2021-02-26
Root cause
Process change control
Distribution
Worldwide distribution - US Nationwide distribution.
Recall number
Z-1334-2021
Recall status
Open, Classified
CDRH record ID
185959
FDA record created
2021-04-01
Recall event number
87396

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗