Open original FDA record ↗Recall record
- FDA identifier
- Z-1335-2012
- Company in FDA record
- Milabs Bv
- FDA classification date
- Apr 5, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
These products have doors which are not equipped with at least one safety interlock which physically disconnects power to the high voltage power supply for the x-ray source. Additionally, these products do not have a certification label and the identification labels do not include the month and year of manufacture.
Source facts
- Action
- A notification will be sent to all affected purchasers by April 15, 2012. The manufacturer will bring the products into compliance with the applicable performance standards within 60 days. Specifically at least one door safety interlock will be replaced per door and revised labels will be produced and applied to the system. These corrections will be made in accordance with MILabs instructions by a MILabs installation and service engineer.
- Initiated
- 2012-04-15
- Root cause
- Radiation Control for Health and Safety Act
- Terminated
- 2012-08-29
- Distribution
- Nationwide Distribution -- Indiana and North Carolina.
- Recall number
- Z-1335-2012
- Recall status
- Terminated
- CDRH record ID
- 108287
- FDA classified
- 2012-04-05
- Recall event number
- 61499