Skip to content

Public FDA record

Z-1336-2015: Boston Scientific: Expel" Drainage Catheter with Twist-Loc" Hub APDL Drainage Catheter and APDL Drainage

Recall record Z-1336-2015 for Boston Scientific Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1336-2015
Company in FDA record
Boston Scientific Corporation
FDA classification date
Mar 27, 2015
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Boston Scientific is initiating this Medical Device Field Correction of the Expel" APD and APDL Drainage Catheters. To date, Boston Scientific has received one complaint for device fragmentation after an Expel" APD catheter was implanted in the biliary system. The most serious foreseeable patient risk for this issue is additional intervention for fragment retrieval using minimally invasive meth

Source facts

Action
Boston Scientific sent an "Urgent Medical Device Correction" letter dated February 25, 2015, to all affected customers. The letter described the problem and the product involved in the recall. The letter informed customers not to use the Expel APD and APDL Drainage catheters for bile drainage and to complete and return the Customer Acknowledgement Form. The letter also noted that the devices are not being retrieved and they are not required to return them to Boston Scientific. For further questions please call (763) 494-7971.
Initiated
2015-02-25
Root cause
Nonconforming Material/Component
Terminated
2016-08-12
Distribution
Worldwide Distribution - US Distribution to the states of : CA, LA, FL, WI, NY, IL, MD and OH., and to the countries of: Belgium, Czech Republic, Finland, France, Germany, Great Britain, Italy, Spain, Switzerland, Austria, Hungary, Netherlands and Portugal.
Recall number
Z-1336-2015
Recall status
Terminated
CDRH record ID
134175
FDA classified
2015-03-27
Recall event number
70632

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.