Open original FDA record ↗Recall record
- FDA identifier
- Z-1336-2015
- Company in FDA record
- Boston Scientific Corporation
- FDA classification date
- Mar 27, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Boston Scientific is initiating this Medical Device Field Correction of the Expel" APD and APDL Drainage Catheters. To date, Boston Scientific has received one complaint for device fragmentation after an Expel" APD catheter was implanted in the biliary system. The most serious foreseeable patient risk for this issue is additional intervention for fragment retrieval using minimally invasive meth
Source facts
- Action
- Boston Scientific sent an "Urgent Medical Device Correction" letter dated February 25, 2015, to all affected customers. The letter described the problem and the product involved in the recall. The letter informed customers not to use the Expel APD and APDL Drainage catheters for bile drainage and to complete and return the Customer Acknowledgement Form. The letter also noted that the devices are not being retrieved and they are not required to return them to Boston Scientific. For further questions please call (763) 494-7971.
- Initiated
- 2015-02-25
- Root cause
- Nonconforming Material/Component
- Terminated
- 2016-08-12
- Distribution
- Worldwide Distribution - US Distribution to the states of : CA, LA, FL, WI, NY, IL, MD and OH., and to the countries of: Belgium, Czech Republic, Finland, France, Germany, Great Britain, Italy, Spain, Switzerland, Austria, Hungary, Netherlands and Portugal.
- Recall number
- Z-1336-2015
- Recall status
- Terminated
- CDRH record ID
- 134175
- FDA classified
- 2015-03-27
- Recall event number
- 70632