Skip to content

Public FDA record

Z-1337-2018: Gyrus ACMI Round Cutting Burrs - to be used with the Diego Elite system

Recall record Z-1337-2018 for Gyrus ACMI, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1337-2018
Company in FDA record
Gyrus ACMI, Inc.
FDA classification date
Mar 20, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

"Chattering" when cutting burrs are used in bone.

Source facts

Action
As a result, OSTA will issue a Field Safety Notice (FSN) to all customers that received variable exposure burrs of 5mm or greater diameter. The FSN will inform them of the concern and will advise to exercise due caution if using the affected device in proximity to sensitive or vital anatomical structures. On September 6, 2017 the firm sent a URGENT: MEDICAL DEVICE FIELD SAFETY NOTICE to the attention of the Surgical Risk Management Department for the following products: MBUR5060FRCV - Burr, 5mm Round Cutting, Variable Exposure, Standard Length, MBUR6060FRCV - Burr, 6mm Round Cutting, Variable Exposure, Standard Length and MBUR7060FRCV - Burr, 7mm Round Cutting, Variable Exposure, Standard Length for All product with an expiration date prior to August 3, 2017 Description of the problem: Olympus has become aware of an issue that requires your attention. This letter pertains to the Olympus Round Variable Cutting Burrs (Burrs) referenced above. Our records indicate that you may have purchased affected burrs. We are in receipt of a number of non-injury-related complaints that has made Olympus aware of a possibility that the burr may chatter when drilling bone. Although this chatter phenomenon can occur with any type or brand of cutting burr, depending on technique and bone density, Olympus has become aware that the larger diameter variable exposure burrs listed above can exhibit more chatter than other similar devices in some situations, particularly when they are extended to longer lengths. After extensive testing, Olympus has determined that our variable exposure burrs of less than 5mm diameter, and all fixed exposure burr offerings, are not affected by this phenomenon. Because of this, Olympus has elected to issue this field safety notice (FSN). Please refer to the enclosure with this letter that cautions of the possibility of the chatter phenomenon in the larger variable offerings. The FSN also recommends that surgeons exercise due caut
Initiated
2017-08-03
Distribution
NM, MI, and TX EU, Australia, Japan, and Hong Kong
Recall number
Z-1337-2018
Recall status
Open, Classified
CDRH record ID
162618
FDA classified
2018-03-20
Recall event number
79548

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗