Open original FDA record ↗Recall record
- FDA identifier
- Z-1338-2012
- Company in FDA record
- Medtronic Neuromodulation
- FDA classification date
- Mar 30, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Medtronic has confirmed that an algorithm used in the Model 8870 application card software has resulted in nine (9) occurrences of an incorrectly displayed Schedule to replace the pump by date. It is estimated that there are more than 140,000 SynchroMed II pump implants worldwide. A patient with a pump reaching EOS prior to replacement may experience the return of underlying symptoms and/or wit
Source facts
- Action
- Medtronic mailed an "Urgent Medical Device Correction " letter addressed to Dear Healthcare Professional on March 12, 2012. The letter described the problem, Background, Clinical Manifestations, Recommendations, and Additional Information. To ensure there is no interruption in therapy, pump replacements should be scheduled before the ERI alarm sounds and the pump reaches its 7-year end of service. Customers were advised to forward the Urgent Medical Device Correction letter to patients whose name appears on their Pump Replacement List. Reimbursement SupportFor assistance with prior authorization, contact Medtronic Reimbursement at the numbers below: Therapy for Severe Spasticity: (800) 292-2903 Pain Therapies: (866) 962-9909 For questions regarding the letter call 800-328-0810.
- Initiated
- 2012-03-12
- Root cause
- Software design
- Terminated
- 2016-04-29
- Distribution
- Nationwide Distribution
- Recall number
- Z-1338-2012
- Recall status
- Terminated
- CDRH record ID
- 107986
- FDA classified
- 2012-03-30
- Recall event number
- 61391