Z-1338-2012: Medtronic, Model 8870, Application Software Card is part of the SynchroMed II Drug Infusion System. The

Recall record Z-1338-2012 for Medtronic Neuromodulation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1338-2012
Company in FDA record
Medtronic Neuromodulation
FDA classification date
Mar 30, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Medtronic has confirmed that an algorithm used in the Model 8870 application card software has resulted in nine (9) occurrences of an incorrectly displayed Schedule to replace the pump by date. It is estimated that there are more than 140,000 SynchroMed II pump implants worldwide. A patient with a pump reaching EOS prior to replacement may experience the return of underlying symptoms and/or wit

Source facts

Action
Medtronic mailed an "Urgent Medical Device Correction " letter addressed to Dear Healthcare Professional on March 12, 2012. The letter described the problem, Background, Clinical Manifestations, Recommendations, and Additional Information. To ensure there is no interruption in therapy, pump replacements should be scheduled before the ERI alarm sounds and the pump reaches its 7-year end of service. Customers were advised to forward the Urgent Medical Device Correction letter to patients whose name appears on their Pump Replacement List. Reimbursement SupportFor assistance with prior authorization, contact Medtronic Reimbursement at the numbers below: Therapy for Severe Spasticity: (800) 292-2903 Pain Therapies: (866) 962-9909 For questions regarding the letter call 800-328-0810.
Initiated
2012-03-12
Root cause
Software design
Terminated
2016-04-29
Distribution
Nationwide Distribution
Recall number
Z-1338-2012
Recall status
Terminated
CDRH record ID
107986
FDA classified
2012-03-30
Recall event number
61391

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗