Open original FDA record ↗Recall record
- FDA identifier
- Z-1338-2020
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA record creation date
- Feb 20, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
A software bug may lead to one image to be assigned to two different patients, which could potentially affect medical diagnosis.
Source facts
- Action
- On January 8, 2020, the firm notified customers of the product issue via an Urgent Medical Device Correction letter. Customers were instructed to ensure that the image displayed after acquisition is a newly acquired image and belongs to the current patient. If the error occurs and a previously acquired image is duplicated, the user should delete the image and contact Siemens Customer Service at 1-800-888-7436 to report the incident. Siemens Healthineers will correct the error with a field software update which will be available beginning January 2020. Your local Siemens Healthineers Service organization will contact you to arrange a date to perform this corrective action when the software update is available for your system. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436.
- Initiated
- 2020-01-08
- Root cause
- Software design
- Terminated
- 2022-06-21
- Distribution
- US Nationwide distribution.
- Recall number
- Z-1338-2020
- Recall status
- Terminated
- CDRH record ID
- 179459
- FDA record created
- 2020-02-20
- Recall event number
- 84815