Open original FDA record ↗Recall record
- FDA identifier
- Z-1338-2021
- Company in FDA record
- Heartware, Inc.
- FDA record creation date
- Apr 6, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class I
Reason stated in the source
There is the potential for the Controller power cables (AC Adapter, DC Adapter, and Battery cables), HVAD Alarm Adapter, and HVAD Monitor data cables to cause damage to the HVAD Controller ports when inserting the cables into the Controller ports, due to wear from misalignment of the cables to the port.
Source facts
- Action
- The firm issued a Medical Device Notice to its consignees on 02/26/2021 by letter delivered via UPS 2-day delivery. The letter explained the issue and directed the physician to review the patient management recommendations on the frequency of of inspecting the HVAD Controller, HVAD Controller power cables, and HVAD Monitor data cables. For questions regarding this material, please contact your Medtronic Field Representative.
- Initiated
- 2021-02-26
- Root cause
- Device Design
- Distribution
- Worldwide
- Recall number
- Z-1338-2021
- Recall status
- Open, Classified
- CDRH record ID
- 186042
- FDA record created
- 2021-04-06
- Recall event number
- 87446