Skip to content

Public FDA record

Z-1338-2023: Moberg Component Neuromonitoring System (CNS Monitor)-Intended as a short-term neurological monitoring sy

Recall record Z-1338-2023 for Moberg Research, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1338-2023
Company in FDA record
Moberg Research, Inc.
FDA classification date
Apr 4, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Battery leakage can cause corrosion to the metal enclosure near the electrical interface. Prolonged contact to corrosive materials could cause injury to skin. DOC

Source facts

Action
Natus issued Letters Urgent Medical Device Recall Letter Component Neuromonitoring System (Moberg CNS Monitor Model CNS-350) via first class beginning Feb 20, 2023. Letter states reason for recall, health risk and action to take: Ensure that the CNS is either shut down or plugged into a power outlet. Perform the next step while using proper personal protective equipment (PPE), including gloves. On the top of the CNS-350 device, open each of the three (3) battery ports and remove any backup battery. 2) Inspect the battery and the battery port for any white, brown, or green/blue buildup, or for any corrosion of the metal contact locations. a. Dispose of the battery at your local electronic e-waste facility. Note: The CNS-350 device may continue to be used (while plugged into a power outlet) with the backup battery removed. b. If battery leakage or corrosion is suspected, cease use of the CNS-350 device. New replacement battery will be shipped to you. The Moberg Technical Service department will be in contact with you in relation to the provision of the replacement product. For new battery replacement, installation instructions will be provided with the battery.
Initiated
2023-02-20
Root cause
Under Investigation by firm
Distribution
Worldwide distribution - US Nationwide and the countries of France, Germany, Italy, Sweden.
Recall number
Z-1338-2023
Recall status
Open, Classified
CDRH record ID
198840
FDA classified
2023-04-04
Recall event number
91782

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗