Open original FDA record ↗Recall record
- FDA identifier
- Z-1339-04
- Company in FDA record
- Celsion Corporation
- FDA classification date
- Aug 11, 2004
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Software controlling a medical device for patient treatment may malfunction and cause thermal injury to patient.
Source facts
- Action
- Celsion Corporation ceased distribution of the device 6/25/04 and implemented a field correction by company representatives at end user facilities on 6/25/04 consisting of computer clock reset. The firm subsequently notified end users on 7/19/04 by certified letter, which included a warning sticker to be placed on the device control screen. The notification and sticker instructed users not to use the device if there is a possibility of the computer clock transitioning through midnight during the procedure.
- Initiated
- 2004-07-19
- Root cause
- Other
- Terminated
- 2005-03-31
- Distribution
- Devices were distributed to 19 medical facilities and physician end users nationwide by the recalling firm''s contract distributor.
- Recall number
- Z-1339-04
- Recall status
- Terminated
- CDRH record ID
- 34258
- FDA classified
- 2004-08-11
- Recall event number
- 29633