Z-1339-04: Prolieve Thermodilatation System, model number M0068808000, labeled in part***Distributed by Boston Scien

Recall record Z-1339-04 for Celsion Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1339-04
Company in FDA record
Celsion Corporation
FDA classification date
Aug 11, 2004
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Software controlling a medical device for patient treatment may malfunction and cause thermal injury to patient.

Source facts

Action
Celsion Corporation ceased distribution of the device 6/25/04 and implemented a field correction by company representatives at end user facilities on 6/25/04 consisting of computer clock reset. The firm subsequently notified end users on 7/19/04 by certified letter, which included a warning sticker to be placed on the device control screen. The notification and sticker instructed users not to use the device if there is a possibility of the computer clock transitioning through midnight during the procedure.
Initiated
2004-07-19
Root cause
Other
Terminated
2005-03-31
Distribution
Devices were distributed to 19 medical facilities and physician end users nationwide by the recalling firm''s contract distributor.
Recall number
Z-1339-04
Recall status
Terminated
CDRH record ID
34258
FDA classified
2004-08-11
Recall event number
29633

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗