Open original FDA record ↗Recall record
- FDA identifier
- Z-1339-2019
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA record creation date
- May 15, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
In affected Artis systems the movement of the floating tabletop may be blocked after a collision sensor has been activated during system movement.
Source facts
- Action
- On April 9, 2019, the firm distributed Urgent Medical Device Correction letters to affected customers. Customers were informed that ARTIS systems equipped with certain motor control units may encounter with hardware failure, and planned procedures may need to be terminated and performed on an alternative x-ray system. The firm stated their service organization would contact customers shortly to arrange a date to perform a software update to correct the issue. For an earlier appointment or if you have any questions, you may contact the firm at 1-800-888-7436. Customers were asked to promptly notify and instruct all staff at the organization who need to be aware of the notice. Additionally, if you further distributed the product, please identify your customers and notify them at once of this recall.
- Initiated
- 2019-04-09
- Root cause
- Software design
- Terminated
- 2020-04-23
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of AL,AR, AZ, CA, FL, ID, IL, IN, KS, LA, MA, MD, MI, MN, MO, NC, ND, NE, NV, NY, OH, OK, OR, PA, RI, SC, SD, TX, UT, VA , WA WI and WV. and Foreign countries.
- Recall number
- Z-1339-2019
- Recall status
- Terminated
- CDRH record ID
- 172575
- FDA record created
- 2019-05-15
- Recall event number
- 82730