Open original FDA record ↗Recall record
- FDA identifier
- Z-1340-2013
- Company in FDA record
- Linvatec Corp. dba ConMed Linvatec
- FDA classification date
- May 20, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The VP1600 HD Still Capture Image System is recalled due to an anomaly between the software code in the VP1600 console and the communications interface for various keyboard key strokes (Caps lock and Num Lock) which may lock up and requires the user to reboot to continue.
Source facts
- Action
- The firm, ConMed, sent an "URGENT: MEDICAL DEVICE RECALL NOTIFICATION" letter dated 3/8/2013 to its customers after initial phone notifications. The letter identified the product, problem, and actions to be taken by the customers.
- Initiated
- 2013-03-18
- Root cause
- Software design (manufacturing process)
- Terminated
- 2013-11-06
- Distribution
- Worldwide Distribution-USA (nationwide) including the Virgin Islands and the states of CA, CT, FL, GA, IN, KY, KS, LA, MD, MI, MN, MO, MS, MT, NE, NC, NJ, NY, OH, OR, PA, SC, TN, TX, UT, VA, and WI, and the countries of Austria, Belgium, Canada, Denmark, Hong Kong and Zambia.
- Recall number
- Z-1340-2013
- Recall status
- Terminated
- CDRH record ID
- 116713
- FDA classified
- 2013-05-20
- Recall event number
- 64647