Z-1340-2020: ARTIS Pheno-Interventional Fluoroscopic X-Ray System Model Number: 10849000 - Product Usage:AXIOM Artis i

Recall record Z-1340-2020 for Siemens Medical Solutions USA, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1340-2020
Company in FDA record
Siemens Medical Solutions USA, Inc
FDA record creation date
Feb 20, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential exists for one or more axes of the robotic stand to become blocked by their safety brakes; In the event the axis is blocked by its safety brake, the robotic stand can no longer perform the intended movements

Source facts

Action
Siemens issued Urgent Medical Device Correction letter dated January 24, 2020 A via AX064/19/S ( corrective action via AX063/19/S), stating reason for recall, health risk and a corrective action: In the event the robotic stand shows stuttering movements or a squealing noise while initiating movements, do not move the robotic stand and C-arm any longer. Siemens advises that an alternative system is available to continue treatment in case of occurrence. Please follow the standard emergency procedures you have in place in the event of system failures. It is recommended to consider reinforcing these emergency procedures until the update has been completed. Siemens will correct the hardware error by replacing all affected relays and altering the wiring via Update Instruction AX063/19/S. Our service organization will contact you shortly to arrange a date to perform this corrective action. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436. Following the correction, the cause is eliminated, and any recurrence of this potential fault is prevented
Initiated
2020-01-24
Root cause
Device Design
Terminated
2020-07-10
Distribution
US Nationwide distribution.
Recall number
Z-1340-2020
Recall status
Terminated
CDRH record ID
179548
FDA record created
2020-02-20
Recall event number
84858

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗