Skip to content

Public FDA record

Z-1340-2026: Brand Name: Leksell GammaPlan Product Name: Radionuclide RT Treatment planning system Software Version: 1

Recall record Z-1340-2026 for Elekta, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1340-2026
Company in FDA record
Elekta, Inc.
FDA classification date
Feb 10, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to an error when defining a new stereotactic reference that users fail to first change an already defined obsolete reference to a pre-plan reference. This may result in a treatment plan that potentially irradiates the incorrect patient location if the error is not detected by users.

Source facts

Action
On 12/12/2025, the firm sent an "IMPORTANT FIELD SAFETY NOTIFICATION/URGENT: MEDICAL DEVICE CORRECTION" letter via email to customers informing them that when Defining a new stereotactic reference in an examination for a frame treatment while failing to first change an already defined obsolete stereotactic reference (representing a non-current frame fitting) to a pre-plan reference, may result in a treatment plan that irradiates the wrong patient location if the resulting inconsistent alignment between the stereotactic references is not detected by the user. Customers are instructed to: If the stereotactic reference for a frame treatment must be changed prior to plan approval (e.g. due to a change of frame fitting), make sure to first change the already defined obsolete stereotactic reference to a pre-plan reference before defining the new stereotactic reference. By so doing, the treatment examination is automatically updated according to the new stereotactic reference when it is defined. Do not combine the use of a stereotactic CBCT reference with other stereotactic references for treatments with a frame. This specific recommendation for CBCT is motivated by the increased likelihood of failure for users to detect an inconsistent stereotactic CBCT reference relative to other stereotactic references, compared to MR and CT image studies, which are typically displayed and explored throughout the planning process. Post this notice in a place accessible to all users, e.g. Instructions for Use, until this action is closed. " Advise the appropriate personnel, working with this product, on the content of this letter. For Questions - Contact Elekta Care Support at 855-693-5358 or email Application.Support.NA@elekta.com
Initiated
2025-12-12
Root cause
Use error
Distribution
Worldwide - U.S. Nationwide distribution including in the states of AK, AZ, CA, CO, ST, FL, GA, HI, IA, IL, LA, MA, MI, MN. MO, NC, NJ, NM, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, WA, WI, and WV. The countries of Algeria, Argentina, Australia, Belarus, Belgium, Brazil, Bulgaria, Canada, China, Colombia, Croatia, Czech Republic, Dominican Republic, Ecuador, Egypt, El Salvador, France, Georgia, Germany, Hong Kong, India, Iran, Iraq, Italy, Japan, Jordan, Republic of Korea, Kuwait, Libyan Arab Jamahiriya, Lithuania, Malaysia, Mexico, Morocco, Netherlands, Norway, Pakistan, Peru, Poland, Romania, Russian Federation, Saudi Arabia, South Africa, Spain, Sweden, Taiwan, Thailand, Turkey, United Kingdom, and Vietnam.
Recall number
Z-1340-2026
Recall status
Open, Classified
CDRH record ID
217904
FDA classified
2026-02-10
Recall event number
98212

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗