Open original FDA record ↗Recall record
- FDA identifier
- Z-1341-2018
- Company in FDA record
- Beckman Coulter Inc.
- FDA record creation date
- Apr 10, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The device may process the same sample with two different sample IDs and sample information while using the single tube loader, which has the potential for erroneous results due to the mis-identification.
Source facts
- Action
- The recalling firm issued letters dated 1/25/2018 via email on 1/31/2018 or to some customers via regular mail during the week of 1/29/2018.
- Initiated
- 2018-01-29
- Root cause
- Software design
- Terminated
- 2019-06-27
- Distribution
- Distribution was made nationwide in the U.S. and in Puerto Rico. There was also government distribution, but no military distribution. Foreign distribution was made to Angola, Australia, Belgium, Botswana, Brazil, Canada, Chile, Colombia, Czech Republic, Denmark, Finland, France, Germany, Ghana, Hong Kong, India, Israel, Italy, Japan, Kazakhstan, Kenya, Lesotho, Malawi, Malaysia, Mexico, Namibia, Netherlands, New Zealand, Norway, Pakistan, Panama, Poland, Portugal, Romania, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, Uganda, Ukraine, United Kingdom, and Zambia.
- Recall number
- Z-1341-2018
- Recall status
- Terminated
- CDRH record ID
- 162185
- FDA record created
- 2018-04-10
- Recall event number
- 79381