Z-1341-2019: Artis Q (material 10848280, 10848281, 10848282) - Product Usage: AXIOM Artis zee and Q/ Q.zen is a family

Recall record Z-1341-2019 for Siemens Medical Solutions USA, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1341-2019
Company in FDA record
Siemens Medical Solutions USA, Inc
FDA record creation date
May 15, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

In affected Artis systems the movement of the floating tabletop may be blocked after a collision sensor has been activated during system movement.

Source facts

Action
On April 9, 2019, the firm distributed Urgent Medical Device Correction letters to affected customers. Customers were informed that ARTIS systems equipped with certain motor control units may encounter with hardware failure, and planned procedures may need to be terminated and performed on an alternative x-ray system. The firm stated their service organization would contact customers shortly to arrange a date to perform a software update to correct the issue. For an earlier appointment or if you have any questions, you may contact the firm at 1-800-888-7436. Customers were asked to promptly notify and instruct all staff at the organization who need to be aware of the notice. Additionally, if you further distributed the product, please identify your customers and notify them at once of this recall.
Initiated
2019-04-09
Root cause
Software design
Terminated
2020-04-23
Distribution
Worldwide distribution - US Nationwide distribution in the states of AL,AR, AZ, CA, FL, ID, IL, IN, KS, LA, MA, MD, MI, MN, MO, NC, ND, NE, NV, NY, OH, OK, OR, PA, RI, SC, SD, TX, UT, VA , WA WI and WV. and Foreign countries.
Recall number
Z-1341-2019
Recall status
Terminated
CDRH record ID
172577
FDA record created
2019-05-15
Recall event number
82730

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗