These 18 probes may have mis-wiring in the inner cables, causing decreased sensitivity and thus resulting in lower peak velocity readings.
Source facts
Action
On January 31, 2020, the firm initiated customer communications via Urgent Medical Device Recall letters. Customers were informed of the potential mis-wiring issue. Clinicians are advised to discontinue use of the probes to minimize adverse events. Hitachi Service will contact each affected site to schedule a service visit. Hitachi will test all affected probes and remove and replace any probes that do not pass the test. The correction is expected to begin February 2020.
Initiated
2020-01-31
Root cause
Process change control
Terminated
2020-07-31
Distribution
Domestic distribution to CA, FL, GA, MA, NV, OH, TX, and WV. Foreign distribution to Canada.
Recall number
Z-1341-2020
Recall status
Terminated
CDRH record ID
179521
FDA record created
2020-02-20
Recall event number
84838