Z-1341-2026: Laerdal Compact Suction Unit 4 RTCA version (LCSU 4 RTCA) 1) LCSU 4, 800 ml, RTCA; Catalog Number: 88

Recall record Z-1341-2026 for Laerdal Medical Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1341-2026
Company in FDA record
Laerdal Medical Corporation
FDA classification date
Feb 11, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

units manufactured in a limited time period may emit electromagnetic noise beyond the acceptable limits for RTCA application specified in RTCA DO-160G Chapter 21, Equipment Category M.

Source facts

Action
On January 13, 2026, Medical Device Recall letters were sent to customers. Actions to be taken: 1. Please review this notice carefully and provide acknowledgment that you have received and understood the contents. If you received this notice from a distributor or other third party, please send your acknowledgment directly to Laerdal using the contact information provided below. You must provide the case number in the footer of this notice in your response. 2. Ensure that this notice is shared with all relevant personnel involved in the storage, maintenance, charging, and use of the LCSU 4 RTCA device. 3. Maintain this information for your records. As part of Laerdal s commitment to customer service, RTCA/DO-160G Section 21 Category M compliant replacement units are being offered, free of charge. To receive a replacement, you must reply within 2 weeks of receipt, with the following information: 1. Confirm acknowledgment of this notice as mentioned above. 2. Confirm serial numbers of your affected units, shipping address, and contact information in your reply. Once validated, Laerdal Medical will arrange for your affected unit(s) to be replaced at no cost. Important: Replacement units (881152) are shipped without a carry bag, battery, external power supply, wire stand, canister, or tubing. Please retain these accessories from your existing unit for continued use. Customers are required to confirm responsibility for disposing of the replaced LCSU 4 RTCA Unit(s) in accordance with applicable environmental regulations. Proof of disposal must be retained and made available upon request. By accepting the replacement, the customer releases Laerdal from any future claims or liabilities related to the old unit. Please maintain awareness of this notice and the related actions for your records. ***Update 03/24/2026***Additional recall letters were sent on or around 2/25/2026 to additional consignees identified by the recalling firm.
Initiated
2026-01-13
Root cause
Nonconforming Material/Component
Distribution
Worldwide distribution - US Nationwide and the countries of Denmark, Sweden, Norway, Canada, Finland, Australia, Spain, France, United Kingdom, Italy, Korea, Germany, Netherlands, Switzerland, New Zealand, Singapore, Malaysia, Austria.
Recall number
Z-1341-2026
Recall status
Open, Classified
CDRH record ID
217469
FDA classified
2026-02-11
Recall event number
98039

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗