The firm received customer complaints associated with the Update Kit CD-ROM not being able to generate a valid calibration after being uploaded, resulting in patient results not being generated.
Source facts
Action
Medical Device Correction letter dated March 24, 2010 was sent to customers. The letter identified the affected product and explained the reason for recall. Customers were asked to destroy Lot No. AA00190 and replace it with the Lot No. AA00838 CD-ROM that is enclosed with the letter. Customers are to complete and fax back the Medical Device Correction Response Form at the number provided.
Initiated
2010-03-24
Root cause
Other
Terminated
2011-03-04
Distribution
Worldwide Distribution -- USA and Italy.
Recall number
Z-1342-2011
Recall status
Terminated
CDRH record ID
93350
FDA classified
2011-02-17
Recall event number
56387