Z-1342-2022: Artis Icono, Interventional Fluoroscopic X-Ray System, Model No. 11327600

Recall record Z-1342-2022 for Siemens Medical Solutions USA, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1342-2022
Company in FDA record
Siemens Medical Solutions USA, Inc
FDA record creation date
Jul 6, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

During system tests, an increased wearing of the Image acquisition system fans has been observed. This could lead to an electrical failure which may cause malfunction of the image system during regular system operation.

Source facts

Action
On May 20, 2022, the firm sent Urgent Medical Device Correction letters to affected customers. Customers were informed that in the event this issue occurs, the system will switch to Bypass fluoroscopy mode. In Bypass fluoroscopy mode a limited imaging functionality (non-subtracted, continuous fluoroscopy with reduced power and without acquisition and storage of images) would remain available. This may result in a situation where it is necessary to cancel clinical treatment or to continue treatment on an alternative system. Please ensure that patient treatment can be continued in other ways if there is any possible danger for the safety of the patient. Customer Instructions: In the event you recognize a high squeaking scratching loud noise out of one of the system cabinets, please contact your service partner for immediate action. It is recommended to avoid performing any critical interventions on the system. Due to the possible risk of electrical shock to the technician, please ensure that your service partner (if not Siemens Healthineers) first contacts Siemens Healthineers' service organization before starting any corrective actions. The firm will replace the affected parts. The service organization will contact customers to arrange a date to perform the corrective action. To schedule an earlier appointment, contact the firm at 1-800-888-7436.
Initiated
2022-05-20
Root cause
Process control
Distribution
Domestic distribution to AL, AR, AZ, CA, CO, CT, DE, FL, ID, IL, IN, KY, LA, MA, MD, MI, MO, MT, NC, NE, NJ, NY, OH, PA, RI, SC, SD, TN, TX, UT, VA, WI and the countries of Armenia, Australia, Austria, Belgium, Canada, China, Denmark, Finland, France, Germany, Greece, Hong Kong, Ireland, Israel, Japan, Kazakhstan, Netherlands, New Zealand, Norway, Oman, Saudi Arabia, Singapore, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, United Arab Emirates, United Kingdom, Vietnam.
Recall number
Z-1342-2022
Recall status
Open, Classified
CDRH record ID
194327
FDA record created
2022-07-06
Recall event number
90401

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗