Open original FDA record ↗Recall record
- FDA identifier
- Z-1342-2026
- Company in FDA record
- Alphatec Spine, Inc.
- FDA classification date
- Feb 11, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Due a design issue where the navigated array connection geometry is incorrect.
Source facts
- Action
- On 01/02/2026, the firm emailed an "URGENT-MEDICAL DEVICE RECALL NOTIFICATION" Letter to customers informing them that Alphatec Spine has been notified of complaints regarding LIF Navigated Osteotome instruments appearing to have incorrect orientation when used with the Medtronic StealthStation". The instrument is correctly identified and displays the correct navigated length; however, the projected image on the Medtronic StealthStation" is rotated by 90. Customers are instructed to: "Review their inventory to determine if any of these affected devices are in their possession. If a device is within their possession, abstain from use and return the device to Alphatec Spine using the attached return label. "Share this notice within their organization or any organization where the affected device(s) has (have) been transferred. "Fill out the last page of this letter to confirm that they have read this notification and have taken all necessary actions, as described in this notification. For any questions/concerns or require assistance with the recall, e-mail QMSCompliance@atecspine.com or call 1-800-922-1356.
- Initiated
- 2026-01-02
- Root cause
- Device Design
- Distribution
- U.S. Nationwide distribution in the states of AZ, CA, FL, IL, IN, MA, MI, MO, NC, NH, NY, TN, TX, and VA.
- Recall number
- Z-1342-2026
- Recall status
- Open, Classified
- CDRH record ID
- 218227
- FDA classified
- 2026-02-11
- Recall event number
- 98251