Open original FDA record ↗Recall record
- FDA identifier
- Z-1343-2020
- Company in FDA record
- Hologic, Inc.
- FDA classification date
- Feb 18, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The recalling firm has received reports from the field of the C-arm on the Selenia Dimensions system seizing or dropping during field repairs and cleaning of the systems.
Source facts
- Action
- On January 21, 2020, the firm notified affected customers of the recall via hand-delivered Urgent Medical Device Correction letter. Customers were informed that the firm had received reports of the C-arm seizing or dropping during field repairs and cleaning of the systems. In order to ensure continued safe use of the system, Hologic sent field service engineers to customer sites on January 21, 2020, to inspect affected systems. Customers were asked to not use the system until the inspection was completed. If you have any questions, please contact Technical Support at breasthealth.support@hologic.com
- Initiated
- 2020-01-21
- Root cause
- Process control
- Terminated
- 2021-08-11
- Distribution
- Domestic distribution to 14 states: AK AZ CA CO CT DC FL GA IA ID IL MA MO NC NH NJ NY OH SD TX VA. Foreign distribution to Italy, United Kingdom, Saudi Arabia, Brazil, Costa Rica, Spain, Canada, Cyprus, Germany, France, Korea, Kuwait, The Netherlands, Oman, Peru, and Russia.
- Recall number
- Z-1343-2020
- Recall status
- Terminated
- CDRH record ID
- 179604
- FDA classified
- 2020-02-18
- Recall event number
- 84882