Open original FDA record ↗Recall record
- FDA identifier
- Z-1345-2020
- Company in FDA record
- Philips North America, LLC
- FDA record creation date
- Feb 21, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
if an AAA battery is inserted in the recorder and a user attempts to start it, or if the recorder is inserted in the docking station, the recorder will display Error: 602 and fail to function for a new patient study.
Source facts
- Action
- Philips Issued Medical Device Correction letter dated January 13, 2020 to U.S. customers, via priority mail. Philips Market Organizations are responsible for distributing the letters outside of the U.S. Instructions to Customer Philips is asking customers to follow the Action to be Taken by Customer/User section of the CIL: Upon receipt of this notification, promptly locate and isolate affected devices. The steps in the Instructions on Clearing Error 602 below may be used to reset your devices. Customers can contact Philips should they choose not to reset the device on their own. For information or support contact: your local Philips representative or call us at 1 -800-722-9377.
- Initiated
- 2020-01-13
- Root cause
- Device Design
- Terminated
- 2023-06-22
- Distribution
- US Nationwide Foreign: Afghanistan Albania Algeria Argentina Armenia Aruba Australia Austria Azerbaijan Bangladesh Belarus Belgium Bermuda Bulgaria Burkina Faso Canada China C¿te D'Ivoire Croatia Curacao Cyprus Czech Republic Egypt France Germany Ghana Gibraltar Greece Guatemala Hong Kong Hungary Iceland India Indonesia Iran Iraq Ireland Israel Italy Japan Jordan Kenya Korea, Republic of Kosovo Kuwait Latvia Lebanon Libya Luxembourg Macao Macedonia Malaysia Maldives Mali Malta Mauritius Mexico Mongolia Myanmar Nepal Netherlands Netherlands Antilles New Zealand Nicaragua Nigeria Norway Oman Pakistan Palestine, State of Panama Peru Philippines Poland Portugal Puerto Rico Qatar Romania Russian Federation Rwanda Saudi Arabia Serbia Singapore Slovakia Slovenia South Africa Spain Sri Lanka Suriname Sweden Switzerland Taiwan Thailand Trinidad and Tobago Turkey Uganda United Arab Emirates United Kingdom Uzbekistan Venezuela Viet Nam
- Recall number
- Z-1345-2020
- Recall status
- Terminated
- CDRH record ID
- 179601
- FDA record created
- 2020-02-21
- Recall event number
- 84879