Z-1345-2022: CELLSEARCH Circulating Tumor Cell Kit (Epithelial)-IVD intended for the enumeration of circulating tumor

Recall record Z-1345-2022 for Menarini Silicon Biosystems. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1345-2022
Company in FDA record
Menarini Silicon Biosystems
FDA record creation date
Jul 7, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

High number of total images/unassigned events including (dual positives) and the potential for false positive results being placed into the image gallery in some patient samples

Source facts

Action
Menarini Silicon Biosystems issued notification telephone Urgent Medical Device Correction Notification CTC Kit Lot S108 to US on 5/20/22. Followed by e-mail direct to consignees for US sites - May 23, 2022 and Coordination with Asia Pacific region and the EU with Local QA and RA from the regions. Letter states reason for recall, health risk and action to take: Discontinue usage of Lot S108 and quarantine the lot in your inventory for disposal. Review patient samples run with Kit Lot S108 which had results that were above the cut-off. If you observe any images with the staining pattern of the false positive images shown below or there is any uncertainty regarding these images impacting the patient's outcome from below the cutoff to above the cut-off, the patient sample should be repeated. Report any previous occurrence of unusually high number of total images/ unassigned events or potential false positive events that have not already been reported to the Customer Technical Support. " Complete and return the enclosed Confirmation of Receipt form no later than May 31, 2022. If you have additional questions about this issue, contact Customer Technical Support at +1-877-837-4339 option 2
Initiated
2022-05-20
Root cause
Process control
Distribution
CA FL IL IN MA MI NC NM NY OH ON PA TX Foreign: Italy, Shanghai, Singapore,
Recall number
Z-1345-2022
Recall status
Open, Classified
CDRH record ID
194354
FDA record created
2022-07-07
Recall event number
90417

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗