Z-1346-2013: cobas 4800 system z480 instrument. Integrates fully automated total nucleic acid isolation directly fr

Recall record Z-1346-2013 for Roche Molecular Systems, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1346-2013
Company in FDA record
Roche Molecular Systems, Inc.
FDA classification date
May 21, 2013
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

In rare instances, channel shifted results have been generated with cobas 4800 assays run on v1.1.1 of the cobas 4800 system. Raw data review indicates that, in these cases, all signals are shifted by one channel: channel 1 results report as channel 2, channel 2 results report as channel 3, channel 3 results report as channel 4, and channel 4 results report as channel 1. In the reported

Source facts

Action
Roche sent an "Urgent Medical Device Correction" (UMDC) notification with response form dated April 1, 2013, to all affected customers via UPS Ground on April 1, 2013. The letter identified the product, the problem, and the action to be taken by the customer. A contact number (Roche Diagnostic Technical Support) is provided for technical support 24 hours a day, seven days a week at phone number: 1-888-912-7090.
Initiated
2013-04-01
Root cause
Software design
Terminated
2014-08-07
Distribution
Nationwide Distribution.
Recall number
Z-1346-2013
Recall status
Terminated
CDRH record ID
117447
FDA classified
2013-05-21
Recall event number
64930

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗