Open original FDA record ↗Recall record
- FDA identifier
- Z-1346-2013
- Company in FDA record
- Roche Molecular Systems, Inc.
- FDA classification date
- May 21, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
In rare instances, channel shifted results have been generated with cobas 4800 assays run on v1.1.1 of the cobas 4800 system. Raw data review indicates that, in these cases, all signals are shifted by one channel: channel 1 results report as channel 2, channel 2 results report as channel 3, channel 3 results report as channel 4, and channel 4 results report as channel 1. In the reported
Source facts
- Action
- Roche sent an "Urgent Medical Device Correction" (UMDC) notification with response form dated April 1, 2013, to all affected customers via UPS Ground on April 1, 2013. The letter identified the product, the problem, and the action to be taken by the customer. A contact number (Roche Diagnostic Technical Support) is provided for technical support 24 hours a day, seven days a week at phone number: 1-888-912-7090.
- Initiated
- 2013-04-01
- Root cause
- Software design
- Terminated
- 2014-08-07
- Distribution
- Nationwide Distribution.
- Recall number
- Z-1346-2013
- Recall status
- Terminated
- CDRH record ID
- 117447
- FDA classified
- 2013-05-21
- Recall event number
- 64930