Open original FDA record ↗Recall record
- FDA identifier
- Z-1346-2020
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA record creation date
- Feb 21, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
A potential safety issue arises when an offset calculation is followed by a filter operation; correcting the image alignment after this sequence will result in incorrect offset values which could lead to incorrect repositioning of the patient and dose to wrong location.
Source facts
- Action
- The firm started notifying customers on January 16, 2020, with letters titled Urgent Medical Device Correction. Customers were advised of the product safety issue. To avoid the possibility of an incorrect alignment of the patient, the user must ensure that filtered images arrive on the COHERENCE Oncologist, or apply any additional filters to the image BEFORE using the "interactive shift" or "landmark registration" tools. The firm will be offering a software solution for the issue free of charge. The firm's letter stated that based on their investigation, customers may continue to use their devices.
- Initiated
- 2020-01-16
- Root cause
- Software Design Change
- Terminated
- 2021-01-04
- Distribution
- Domestic distribution to CA, FL, GA, IA, MA, MI, NE, NJ, NV, NY, OH, OK, PA, SD, TN, WV. Worldwide foreign distribution.
- Recall number
- Z-1346-2020
- Recall status
- Terminated
- CDRH record ID
- 179419
- FDA record created
- 2020-02-21
- Recall event number
- 84795