Z-1346-2021: Ysio X.pree, Model Number: 11107464 - Product Usage: A Digital X-Ray System generate Xray images from th

Recall record Z-1346-2021 for Siemens Medical Solutions USA, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1346-2021
Company in FDA record
Siemens Medical Solutions USA, Inc
FDA record creation date
Apr 2, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A patient image was assigned to a different patient during image recovery process. In the unlikely event of interrupted connection to a MAX detector during initiated image recovery, instead of the currently acquired image, previously acquired image (potentially image of the previous patient) may be displayed to the user, potentially impact diagnosis of the patient examined when the connection was interrupted

Source facts

Action
Siemens Medical Solutions USA, Inc. issued Customer Safety Advisory Notice to customers via XP051/20/S on 3/10/21. Letter states reason for recall, health risk and action to take: It is advised for the user to check the image displayed following image recovery to ensure that it is a newly acquired image of the current patient. If the operator detects the error, the image should be deleted, and Siemens Customer Service should be notified. Siemens Healthineers will provide a field software update VA10D for YSIO X.pree units that will eliminate the issue described. The software update will be disturbed to all potentially affected users via an Update Instruction XP052/20/S. Apart from the described correction, the updated software contains further stability, usability, and performance improvements. This resolution will be provided free of charge and is planned will be available in the second quarter of calendar year 2021. Your local Siemens Healthineers Service organization will contact you as soon as the software update is available for your system. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436. Following the corrective action, the cause is eliminated and any recurrence of this potential fault is prevented. Acknowledge Receipt of this Safety Advisory Notice. Additionally, a software solution VA10D to eliminate the root cause of this problem will be distributed via XP052/20/S. Following the corrective action, the cause is eliminated and a recurrence of the fault is prevented.
Initiated
2021-03-10
Root cause
Device Design
Terminated
2023-12-11
Distribution
US Nationwide distribution in the states of FL, MN, OH, PA, TX.
Recall number
Z-1346-2021
Recall status
Terminated
CDRH record ID
186280
FDA record created
2021-04-02
Recall event number
87563

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗