Skip to content

Public FDA record

Z-1346-2023: CAPIOX NX19 Hollow Fiber Oxygenator, Catalog Number 3CX*NX19RE

Recall record Z-1346-2023 for Terumo Cardiovascular Systems Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1346-2023
Company in FDA record
Terumo Cardiovascular Systems Corporation
FDA classification date
Apr 5, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Some Capiox NX, RX and FX Oxygenators may contain an arterial thermistor component that is improperly bonded to the oxygenator port. This may result in a leak at the point where the arterial thermistor is improperly bonded to the oxygenator port.

Source facts

Action
An URGENT MEDICAL DEVICE RECALL notification letter dated 3/1/23 was sent to customers. RECOMMENDED ACTIONS Terumo CVS advises users to remove all affected product from inventory and return unused product to Terumo CVS following the instructions provided in this notice. Terumo CVS will credit the user s accounts for returned, unused products. AFFECTED POPULATION The affected population is all patients on which an impacted Capiox NX, RX or FX Oxygenator is used. CUSTOMER INSTRUCTIONS - Review this Medical Device Recall notice and ensure that all users receive notice of this issue. Refer to the Customer Response Form to identify your product inventory that is subject to this action. Confirm receipt of this communication by completing and returning the attached Customer Response Form to the email address or fax number indicated on the form. Please return the Customer Response Form even if you no longer have the affected product in inventory. If the affected product has been distributed to another user or location, please forward a copy of the notice as needed. Terumo CVS will issue a Return Goods Authorization upon receipt of the Customer Response Form. Return all affected products to Terumo CVS. QUESTIONS? We encourage you to contact Terumo CVS with any questions or concerns: Terumo CVS Customer Service: 1.800.521.2818 Monday Friday, 8 a.m. 6 p.m. ET Terumo Recall Fax: 443-993-5738 or emailing 4439935738@efax.com Terumo Recall Email: TCVSCustomerResponseElkton@terumomedical.com
Initiated
2023-03-01
Root cause
Process control
Distribution
US Nationwide. Belgium, Canada, Costa Rica, Thailand.
Recall number
Z-1346-2023
Recall status
Open, Classified
CDRH record ID
198830
FDA classified
2023-04-05
Recall event number
91775

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗