Open original FDA record ↗Recall record
- FDA identifier
- Z-1347-2016
- Company in FDA record
- GE Medical Systems, LLC
- FDA record creation date
- Apr 7, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Ferrous material in the SIGNA Pioneer Table Control Box (TCB) could cause unexpected attraction to the MR system's high strength magnetic field during servicing activities. This could create a potential risk of serious injury to service personnel who attempt to inappropriately remove the TCB from the magnetic field. There have been no events reported as a result of this issue. Clinical scanni
Source facts
- Action
- Consignees were sent a GE Healthcare "Urgent Medical Device Correction" letter GEHC Ref# 68002 dated March 1, 2016. The letter was addressed to Hospital Administrators / Risk Managers, Radiology Department Managers, and Radiologists. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction and Contact Information. For questions or concerns regarding this notification,contact GE Healthcare Service at 18004371171 or your local Service Representative.
- Initiated
- 2016-03-01
- Root cause
- Labeling Change Control
- Terminated
- 2016-08-16
- Distribution
- Worldwide Distribution - US Nationwide in the states of SC, NC and the countries of Australia, Canada, France. Germany, Japan and Turkey.
- Recall number
- Z-1347-2016
- Recall status
- Terminated
- CDRH record ID
- 144455
- FDA record created
- 2016-04-07
- Recall event number
- 73555