Open original FDA record ↗Recall record
- FDA identifier
- Z-1347-2020
- Company in FDA record
- Zimmer Biomet, Inc.
- FDA classification date
- Jan 29, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Potential increase in fracture rate was detected during internal testing, in addition to potential discoloration inside the inner pouch of the packaging.
Source facts
- Action
- On January 7, 2020, the firm distributed Urgent Medical Device Recall letters to affected customers. Customers were instructed to take the following actions: 1. Review the recall information and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. The sales representative will remove the affected product from your facility. 3. There are no specific patient monitoring instructions related to this recall that are recommended beyond the existing follow-up schedule. IF there is a device fracture, it can be detected through x-rays. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to CorporateQuality.PostMarket@zimmerbiomet.com.
- Initiated
- 2020-01-07
- Root cause
- Under Investigation by firm
- Terminated
- 2021-04-08
- Distribution
- Domestic distribution to AL CA FL IA IL MD MI MN MO MS NC NE NJ NY PA TX UT VA WA WI
- Recall number
- Z-1347-2020
- Recall status
- Terminated
- CDRH record ID
- 179265
- FDA classified
- 2020-01-29
- Recall event number
- 84753