Open original FDA record ↗Recall record
- FDA identifier
- Z-1348-2009
- Company in FDA record
- Synthes Spine
- FDA classification date
- Jun 10, 2009
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Tip breakage, due to inadequate instructions for use. The Technique Guide provided incorrect information related to use of the Torque Limiting Attachment during screw insertion and final tightening.
Source facts
- Action
- All affected Synthes Spine Sales Consultants were notified of the Labeling Correction via email on February 10, 2009. They were instructed to review changes with their surgeons. They were supplied with an electronic version of the corrected Technique Guide on March 11, 2009. All affected Synthes Sales Consultants and Customers were notified of this labeling change via distribution of the updated Technique Guide and a notification letter. This packet was sent via USPS certified mail on March 26, 2009. To ensure a broad and accurate distribution of notification, affected consultants and customers are defined as having been shipped Zero-P 3.0 mm Titanium Cervical Spine Locking Screws. These devices are used only with the Zero-P system.
- Initiated
- 2009-02-10
- Root cause
- Labeling design
- Terminated
- 2009-10-05
- Distribution
- Distributed Nationwide and Canada.
- Recall number
- Z-1348-2009
- Recall status
- Terminated
- CDRH record ID
- 81257
- FDA classified
- 2009-06-10
- Recall event number
- 51781