Power Pack (component) was shipped with an internal battery pack tested to UN-38.3, UL-2054 and IEC-62133 (2nd Edition) standards; however, not tested to IEC-60601-1-2 (3rd edition) standards
Source facts
Action
Email was sent out to healthcare practitioners regarding the recall on September 26, 2019.
Please courier your TLC-2000 to Theralase, at your cost, for modification. A Field Inspection Test will be completed and once passed, a Special Inspections for Medical Electrical Equipment and Systems label will be applied to your TLC-2000 Power Pack..
Initiated
2019-09-26
Root cause
Process control
Terminated
2020-12-03
Distribution
U.S.A.; Australia, Bahrain, Canada, Hong Kong Province of China, Jordan, Mexico, Pakistan, Philippines, Thailand, Romania
Recall number
Z-1348-2020
Recall status
Terminated
CDRH record ID
179526
FDA classified
2020-02-13
Recall event number
84842