Open original FDA record ↗Recall record
- FDA identifier
- Z-1349-2016
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA classification date
- Apr 7, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
To inform users about the possible incorrect values for Distance Measurements when using certain modalities in combination with syngo Imaging.
Source facts
- Action
- Siemens mailed a Customer Advisory Notice to their customers instructing them to check their modalities for the usage of the mentioned DICOM attributes with different values. They recommended upgrading the workplace(s) to syngo Imaging Studio version VB36D if they have such modalities and it cannot be ensured by the modality provider that only one attribute is filled or both attributes are filled with the same value. Customers were asked to contact their local service organization to clarify the compatibility of their syngo Imaging server version with workplace version VB36D.
- Initiated
- 2016-01-18
- Root cause
- Software design
- Terminated
- 2017-08-03
- Distribution
- US (nationwide) including the states of AL, CA, CO, CT, FL, IA, IN, KS, LA, MA, MD, MI, MN, MO, MS, NC, NE, NJ, NY, OK, OR, PA, TN, TX, and WI.
- Recall number
- Z-1349-2016
- Recall status
- Terminated
- CDRH record ID
- 144261
- FDA classified
- 2016-04-07
- Recall event number
- 73475