Open original FDA record ↗Recall record
- FDA identifier
- Z-1349-2018
- Company in FDA record
- GETINGE US SALES LLC
- FDA classification date
- Apr 10, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The manufacturer received complaints indicating that the device's central handle holder detaches easily from the handle holder ring on the light head during clinical use. In the event of such detachment, particles from the handle holder ring could fall into the sterile field during a procedure.
Source facts
- Action
- The firm, Getinge, sent an "URGENT MEDICAL DEVICE RECALL" Notice dated 2/7/2018 to all affected consignees via mail. The notice described the product, problem and actions to be taken. The consignees were instructed to do the following: 1) Ensure that all users of the device are made aware of this field correction. 2) Check the device serial number against the list of affected serial numbers. 3) Contact the recalling firm to arrange for servicing of all affected units. Getinge Customer Service: (800) 950-9912, 8AM - 5PM (EST) Monday to Friday. 4) Complete and sign the enclosed customer response form, and return it by fax to: (973) 860-1675 or by email to: VolistaLights.FA@getinge.com. If you have any questions, please contact your Getinge representative or call Getinge Customer Service at (800)950-9912 (Select option 1) between 8:00 AM and 5:00 PM (EST).
- Initiated
- 2018-02-07
- Root cause
- Component design/selection
- Terminated
- 2021-09-02
- Distribution
- Worldwide Distribution: US (nationwide) to states of: AR, AZ, CA, CO, FL, ID, IA, KS, KY, ME, MO, NE, NY, OR, TN, TX, UT, and WA; and countries of: Argentina, Australia, Austria, Belgium, Bolivia, Brazil, Cameroon, Canada, Chile, China, Colombia, Croatia, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Finland, France, Germany, Ghana, Hong Kong, Hungary, India, Iran, Iraq, Italy, Japan, Jordan, Kenya, Latvia, Lebanon, Luxembourg, Malaysia, Mexico, Namibia, Netherlands, Nigeria, Oman, Pakistan, Panama, Paraguay, Poland, Portugal, South Korea, Romania, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Tanzania, Thailand, Trinidad-Tobago, Turkey, UAE, UK, Yemen and Zimbabwe.
- Recall number
- Z-1349-2018
- Recall status
- Terminated
- CDRH record ID
- 162696
- FDA classified
- 2018-04-10
- Recall event number
- 79583