Z-1350-2020: Ulrich medical neon3 OCT spinal stabilization, REF CS 3917-12, Locking screw for crosslink, screw-to-scre

Recall record Z-1350-2020 for ulrich medical USA Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1350-2020
Company in FDA record
ulrich medical USA Inc
FDA record creation date
Mar 2, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to a production error, the weld seam of the "Locking screw for crosslink, screw-to-screw" cannot reliably transmit the required torque of 3.5 Nm during installation.

Source facts

Action
Ulrich notified initially notified customers, via phone call, on 12/20/2019. On about 12/31/2019, customers were also sent an official "URGENT SAFETY INFORMATION: MEDICAL DEVICE CORRECTIONS & REMOVALS" letter. Customers were informed that the "Locking screw for crosslink, screw-to-screw" cannot reliably transmit the required torque of 3.5 Nm during installation. Instructions included cease use of the affected devices, return them to Ulrich, and complete and return the provided "Product Recall" form. Customers were also instructed to inform all users of the devices within each organization are informed. Questions or concerns can be directed to Mr. Louis Milos, Sr. Manager Business and Quality Systems, at 636-519-0268 ext. 105, l.milos@ulrichmedicalusa.com, Monday through Friday 8:00am to 5:00pm CST.
Initiated
2019-12-31
Root cause
Process control
Terminated
2025-11-17
Distribution
AZ, TX, GA, FL, AR, CA
Recall number
Z-1350-2020
Recall status
Terminated
CDRH record ID
178852
FDA record created
2020-03-02
Recall event number
84638

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗