- FDA identifier
- Z-1350-2020
- Company in FDA record
- ulrich medical USA Inc
- FDA record creation date
- Mar 2, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Due to a production error, the weld seam of the "Locking screw for crosslink, screw-to-screw" cannot reliably transmit the required torque of 3.5 Nm during installation.
Source facts
- Action
- Ulrich notified initially notified customers, via phone call, on 12/20/2019. On about 12/31/2019, customers were also sent an official "URGENT SAFETY INFORMATION: MEDICAL DEVICE CORRECTIONS & REMOVALS" letter. Customers were informed that the "Locking screw for crosslink, screw-to-screw" cannot reliably transmit the required torque of 3.5 Nm during installation. Instructions included cease use of the affected devices, return them to Ulrich, and complete and return the provided "Product Recall" form. Customers were also instructed to inform all users of the devices within each organization are informed. Questions or concerns can be directed to Mr. Louis Milos, Sr. Manager Business and Quality Systems, at 636-519-0268 ext. 105, l.milos@ulrichmedicalusa.com, Monday through Friday 8:00am to 5:00pm CST.
- Initiated
- 2019-12-31
- Root cause
- Process control
- Terminated
- 2025-11-17
- Distribution
- AZ, TX, GA, FL, AR, CA
- Recall number
- Z-1350-2020
- Recall status
- Terminated
- CDRH record ID
- 178852
- FDA record created
- 2020-03-02
- Recall event number
- 84638