Church & Dwight First Response Pregnancy Kits was marketed with the Easy Read App without a new 510k. There is also a potential for Easy Read App to be unable to reads the test results (i.e., no results reported on the APP screen) or to misread the analog results shown on the pregnancy test stick and may lead to misinterpretation of false positive or false negative tests by the user.
Source facts
Action
Church & Dwight Company Easy Read App requested the developer to remove the App from the Apple and Google Play app stores on 1/25/23 and third-party vendor disabled the App on consumers devices. Consumers who try to use the downloaded App receive a message that the App is no longer available.
Initiated
2023-01-25
Root cause
No Marketing Application
Distribution
US Nationwide distribution.
Recall number
Z-1350-2023
Recall status
Open, Classified
CDRH record ID
198788
FDA classified
2023-04-07
Recall event number
91758
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.