Z-1350-2023: Easy Read App associated with First Response Pregnancy Test Sticks - Intended use of converting analog p

Recall record Z-1350-2023 for Church & Dwight Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1350-2023
Company in FDA record
Church & Dwight Inc
FDA classification date
Apr 7, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Church & Dwight First Response Pregnancy Kits was marketed with the Easy Read App without a new 510k. There is also a potential for Easy Read App to be unable to reads the test results (i.e., no results reported on the APP screen) or to misread the analog results shown on the pregnancy test stick and may lead to misinterpretation of false positive or false negative tests by the user.

Source facts

Action
Church & Dwight Company Easy Read App requested the developer to remove the App from the Apple and Google Play app stores on 1/25/23 and third-party vendor disabled the App on consumers devices. Consumers who try to use the downloaded App receive a message that the App is no longer available.
Initiated
2023-01-25
Root cause
No Marketing Application
Distribution
US Nationwide distribution.
Recall number
Z-1350-2023
Recall status
Open, Classified
CDRH record ID
198788
FDA classified
2023-04-07
Recall event number
91758

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗