Z-1351-2022: Philips Laser System, REF LAS-100, Intermittent Operation, Duty cycle dependent upon device in use 100 -

Recall record Z-1351-2022 for Spectranetics Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1351-2022
Company in FDA record
Spectranetics Corporation
FDA record creation date
Jul 11, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The LAS-100 Laser system may detect an inoperable hardware component during power up, which results in an error code and the system not being operable until code is cleared.

Source facts

Action
On June 08, 2022 Spectranetics issued a "Urgent Medical Device Correction" notification to affected consignees via USPS. The recall notification asked consignees to take the following actions: 1. Philips internal testing confirms that 1-3 reboots of the system will temporarily resolve the issue. Philips also recommends the continued use of the Philips Laser System, and to follow the Operator s Manual, as there is no expected harm to the user or patient. Philips recommends notifying all Philips Laser System users within your facility of this communication and retain a copy of this letter for reference. To acknowledge receipt of this notification, please complete, sign, and return the Customer Reply Form, included with this letter, within 30 days upon receipt of this notice to the following Email: igtdc.r@philips.com If you are a distributor or have forwarded affected device(s) to another end user, it is imperative that all end-users with affected devices receive this URGENT Medical Device Correction. Therefore, send a copy of this notification to any customer to whom you have distributed the affected product. If you need any further information or support concerning this issue, please contact your local Philips representative or Philips IGTD Customer Service: Philips Laser System (PLS) Customer Service: Phone: 1-800-231-0978 Email: laserserviceops@philips.com Hours of Operation: Monday - Friday 9:00AM 5:00PM MST
Initiated
2022-06-01
Root cause
Device Design
Distribution
US: AR AZ CA CO CT FL GA HI IA IL IN KS LA MA MD ME MI MN MO NC NJ NY OK PA TN TX VA OUS: Saudi Arabia, Denmark, France, Greece, India, Italy, Netherlands, Serbia, Spain, Sweden, United Kingdom
Recall number
Z-1351-2022
Recall status
Open, Classified
CDRH record ID
194234
FDA record created
2022-07-11
Recall event number
90364

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗